Single Channel LTI-904 Leak Test




The LTI-904 leak test instrument is a one or two channel test system that operates on clean dry shop air to accurately test for product leaks.
TEST METHODS
MASS FLOW  
The LTI-904 displays corrected airflow readings through the use of a mass flow transducer. This state-of-the-art technology provides test results, which are independent of daily changes in barometric pressure and ambient temperature. This leak test instrument measures the leakage or flow of air through the part under test at a specific test pressure. When the leak test is initiated this instrument automatically pressurizes the part, allows the air pressure within the part to stabilize, and measures the actual leakage through the part. The leakage or flow is displayed in actual engineering units.
PRESSURE DECAY/CHANGE  
Two pressure decay/change methods are available depending upon the specific application. Applications with a high leak rate specification or a small test volume utilize a gage-pressure-decay test method. Applications with a low leak rate specification or a large test volume will use a pressure-differential-decay test method. With either method this leak test instrument pressurizes the test part, allows the air pressure within the part to stabilize, and measures the decay/change in pressure.
SETUP & OPERATION
The LTI-904 leak test instrument is easy to setup and operate. The setup procedure is menu driven in response to operator inputs to the front-mounted touch screen display. A maximum of six (6) limit tables for each channel are included with this unit. Setup parameters for these tables include fast fill pressure and tolerance, test pressure and tolerance, cycle times, accept/reject parameters, consecutive reject set point, orifice qualification set points, and various feature flags. Transducer calibration is also menu driven.
SYSTEM ENCLOSURE
The LTI-904 is housed in multiple enclosures linked by an electrical cable(s). One (1) electrical enclosure and up to two (2) pneumatic enclosures are provided. The electrical enclosure contains the Allen Bradley SLC 5/04 processor, Allen Bradley PanelView touch screen display, power supplies, push buttons, and indicator lights. Ports, internal to the enclosure, are available for one (1) data highway, one (1) RS232 printer port, and one (1) optional DB9 for a bar code reader. The pneumatic enclosure(s) contain the pressure and leak transducers, regulators, and valves. Ports are provided on each enclosure for air and power connections. The standard color for the LTI-904 is white.
DIMENSIONS
Control/Pneumatic Enclosure
Pneumatic Enclosure
18" high x 18" wide x 10" deep
16" high x 18" wide x 10" deep
WEIGHT
Control Enclosure
Pneumatic Enclosure
50 pounds
25 pounds
ENVIRONMENTAL CONDITIONS
Operating Temperature
Storage Temperature
Test Part
Relative Humidity
35-130°F (2-55°C)
35-150°F (2-66°C)
Room Temperature
0-95% Non-Condensing
POWER REQUIREMENTS
Input Power (Selectable Internally)
Frequency
Current
105-120VAC or 220-240VAC
50 or 60 Hz
3 amps maximum
SHOP AIR REQUIREMENTS (CLEAN, NON-LUBRICATED, DRY AIR)
Pressure Test Test Input Pressure
125 PSIG maximum (STD)
  Test Pressure
90 PSIG maximum (STD)
Vacuum Test Input Vacuum
30 in Hg maximum (STD)
  Test Vacuum
25 in Hg maximum (STD)
CONTROL INTERFACE
Inputs
(3) BCD Remote Start/Stop per Channel
   
115 VAC or 24 VDC
Outputs (1) End of Test Signal per channel
115 VAC or 24 VDC
  (1) Accept per Channel
115 VAC or 24 VDC
Printer (1) Serial Port Available
Data Output (ASCII)
Host Computer (1) Data Highway
Plant Communications
Barcode (1) Serial Port Available
Data Collection

PERFORMANCE (FLOW RANGE AND SCALING)
Minimum Full Scale Range
Standard Full Scale Range
Maximum Full Scale Range
 
2.00 SCCM
10 SCCM to 30 SLPM
Up to 300 SLPM (Consult Factory)
Accuracy
Linearity
Repeatability
 
Plus or Minus 1% of Full Scale
0.5% of Full Scale Range
Plus or Minus 0.2% of Full Scale
System Warm Up Time
(from Room Temperature)
 
Casual Operation: 5 seconds
Precise Measurement: 10 minutes