LTI-01
 
  • Pressure Decay/Change
  • Pressure or Vacuum
  • Affordable
  • Accuracy to ± .5% Full Scale
  • Easy Setup and Operation
  • 3 Limit Tables

LTI-01 Pressure Decay Test Instrument

The LTI-01 leak test instrument is a single channel test system that operates on clean dry shop air to accurately test for product leaks.
TEST METHODS

PRESSURE DECAY/CHANGE  
This unit is tailored for lower volume applications and utilizes a gage-pressure-decay test method. This leak test instrument pressurizes the test part, allows the air pressure within the part to stabilize and measures the decay/change in pressure over time.
SETUP & OPERATION  
The LTI-01 leak test instrument is easy to setup and operate. The setup procedure is menu driven in response to operator inputs to the front-mounted screen display. Three (3) limit tables are included with this unit. Setup parameters for these tables include fill pressure and tolerance, test pressure and tolerance, cycle times, and accept/reject parameters. Consecutive reject set point and various feature flags are also included in the unit to customize it’s operation. Unit calibration is also menu driven.
SYSTEM ENCLOSURE
The LTI-01 leak test instrument is easy to setup and operate. The setup procedure is menu driven in response to operator inputs to the front-mounted screen display. Three (3) limit tables are included with this unit. Setup parameters for these tables include fill pressure and tolerance, test pressure and tolerance, cycle times, and accept/reject parameters. Consecutive reject set point and various feature flags are also included in the unit to customize it’s operation. Unit calibration is also menu driven.
DIMENSIONS
Control/Pneumatic Enclosure
12” high x 12” wide x 6” deep
WEIGHT
Control/Pneumatic Enclosure
25 pounds
ENVIRONMENTAL CONDITIONS
Operating Temperature
Storage Temperature
Test Part
Relative Humidity
35-130°F (2-55°C)
35-150°F (2-66°C)
Room Temperature
0-95% Non-Condensing
POWER REQUIREMENTS
Input Power
Frequency
Current
115VAC
60 Hz
2.5 amps maximum
SHOP AIR REQUIREMENTS (CLEAN, NON-LUBRICATED, DRY AIR)
Pressure Test Input Pressure
125 PSIG maximum (STD)
  Pressure Test
100 PSIG maximum (STD)
Vacuum Test Input Vacuum
30 in Hg maximum (STD)
  Test Vacuum
25 in Hg maximum (STD)
CONTROL INTERFACE
Inputs
(2) BCD Remote Start/Stop per Channel
24 VDC
Outputs (1) End of Test
24 VDC
  (1) Remote Pass
24 VDC
(1) Remote Fail
24 VDC
Printer (1) RS232 Serial Port Available (Port 2)
Data Output (ASCII Format)
PERFORMANCE (FLOW RANGE AND SCALING)
Standard Full Scale Range
Accuracy
Repeatability
0-15, 30, 60, and 100 PSIG
Plus or Minus .5% of Full Scale
Plus or Minus .25% of Full Scale
System Warm Up Time
(from Room Temperature)

Casual Operation: 5 seconds
Precise Measurement: 5 minutes